Document Type : original article
Authors
1 Post Graduate, Department of Paediatrics, Sri ManakulaVinayagar Medical College and Hospital, Puducherry-605107, India.
2 Associate Professor, Department of Paediatrics, Sri ManakulaVinayagar Medical College and Hospital, Puducherry-605107, India.
3 Professor, Department of Paediatrics, Sri ManakulaVinayagar Medical College and Hospital, Puducherry-605107, India.
Abstract
Background: This study attempted to compare radiographic findings with that of the ultrasound in locating the central line tip and evaluate the agreement between these two diagnostic modalities.
Methods: This hospital-based Prospective Observational Comparative Diagnostic Accuracy Test study was conducted on 51 neonates, in the NICU of a tertiary health care centre. A consecutive sampling technique was used for the neonates included in this study. The central catheter (umbilical vein catheter / peripherally inserted central catheter) was inserted. After the central line insertion, a thoracoabdominal X-ray was done to identify the catheter tip position as the standard protocol. USG assessment of the central line tip position was also done.
Results: Out of the 51 central lines studied, 49 were UVC and 2 were PICC line insertion. USG was able to identify all the malpositioned (n=38) and optimally placed central line tips (n=10) as in X-ray and also additionally identified the three doubtful tips on X-ray as malpositioned (n=1) and optimally placed (n=2). X-ray showed that 78.5% of UVCs were low lying and 66.6% were in intrahepatic IVC as identified by USG. Among the procedure-related complications, post X-ray more complications were noted, the most common being hypothermia.
Conclusion: Our study results show that Point Of Care Ultrasound is a safer, more efficacious, and quick bedside diagnostic test compared to the conventional radiography in detecting central line tip position in neonates, helping pediatricians to quickly confirm the central line tip position before starting infusions.
Keywords
- Thakur A, Kumar V, Modi M, Kler N, Garg P. Use of Point of Care Ultrasound for Confirming Central Line Tip Position in Neonates. Indian Pediatrics. 2020 Sep; 57:805-7.
- Oppenheimer DA, Carroll BA, Garth KE, Parker BR. Sonographic localization of neonatal umbilical catheters. American Journal of Roentgenology. 1982 Jun 1; 138(6):1025-32.
- Simanovsky N, Ofek-Shlomai N, Rozovsky K, Ergaz-Shaltiel Z, Hiller N, Bar-Oz B. Umbilical venous catheter position: evaluation by ultrasound. European radiology. 2011 Sep; 21:1882-6.
- Greenberg M, Movahed H, Peterson B, Bejar R. Placement of umbilical venous catheters with use of bedside real-time ultrasonography. The Journal of pediatrics. 1995 Apr 1; 126(4):633-5.
- Ohki Y, Tabata M, Kuwashima M, Takeuchi H, Nako Y, Morikawa A. Ultrasonographic detection of very thin percutaneous central venous catheter in neonates. Acta paediatrica. 2000 Nov; 89(11):1381-4.
- Madar RJ, Deshpande SA. Reappraisal of ultrasound imaging of neonatal intravascular catheters. Archives of Disease in Childhood-Fetal and Neonatal Edition. 1996 Jul 1; 75(1):62-4.
- Lim-Dunham JE, Vade A, Capitano HN, Muraskas J. Characteristic sonographic findings of hepatic erosion by umbilical vein catheters. J Ultrasound Med 2007; 26:661-6.
- Lam HS, Li AM, Chu WC, Yeung CK, Fok TF, Ng PC. Mal-positioned umbilical venous catheter causing liver abscess in a preterm infant. Biol Neonate 2005; 88:54-6.
- Goh SS, Kan SY, Bharadwaj S, Poon WB. A review of umbilical venous catheter-related complications at a tertiary neonatal unit in Singapore. Singapore medical journal. 2021 Jan; 62(1):29.
- Khaladkar SM, Gujarathi A, Kulkarni VM, Singh A. Hepatic necrosis due to the umbilical vein catheter malposition: A case report with review of literature. Medical Journal of Dr. DY Patil University. 2014 Nov 1; 7(6):793.
- Telang N, Sharma D, Pratap OT, Kandraju H, Murki S. Use of real-time ultrasound for locating tip position in neonates undergoing peripherally inserted central catheter insertion: a pilot study. The Indian journal of medical research. 2017 Mar; 145(3):373.
- Shukla H, Ferrara A. Rapid estimation of insertional length of umbilical catheters in newborns. Am J Dis Child. 1986; 140:786–8.
- El-Maadawy SM, El-Atawi KM, Elhalik MS. Role of bedside ultrasound in determining the position of umbilical venous catheters. Journal of Clinical Neonatology. 2015 Jul 1; 4(3):173.